AD-4530
Digital Indicator for Strain Gauge Sensors
SC series Suitable for measurement of load, pressure,
torque, tension, and more
 
 
PDFBrochure (PDF 646KB)
 
PDFManual (PDF 818KB)
PDFSimplified manual (PDF 288KB)
 
 
 
Please send your inquiry here
Features
Low cost digital indicator for various applications. Suitable for measurement of load, pressure, torque, tension
    and more at an affordable price.
Digital Zero (DZ) stores zero data at any measurement point. This function can also be used to tare weight.
    Measurement can simply be restarted when it was interrupted by a loss of power.
Power on zero function adjusts to zero when the power is turned on.
Digital filter provides stable measurement results.
Digital span calibration. Easy calibration just by inputting the rated output of a sensor. No need to use a mass.
Unit characters - Up to five alphanumeric characters can be added to serial output.
Comparator function displays and outputs HI, OK and LO according to upper and lower limit settings. Hysterisis
    width and time can be set to prevent chattering on the output terminal. This reduces load on the relay circuit.
Hold function. Selectable from sample hold, peak hold, bottom hold and bipolar peak hold. The peak hold
    function allows the user to capture the peak value and easily obtain the maximum value of a destruction test
    or tightening torque.
Latch function holds display value/comparator (optional)/analog output (optional)/serial output (optional)
    corresponding to external latch input.
Various data output (optional). Necessary data can be easily obtained using optional comparator relay output,
    serial output (RS-485/RS-232C), D/A voltage output and D/A current output.
Options
AD4530-200 - Relay output
AD4530-030 - RS485
AD4530-040 - RS-232C input/output
AD4530-007 - D/A analog output
AD4530-237 - Relay output, RS485 and D/A analog output
AD4530-247 - Relay output, RS232C and D/A analog output
*The options above cannot be used at the same time.
Specifications
Measurement points 1
Sensor type Strain gauge sensors
Sensor power supply 5 V, 3 sensors of 350 Ω maximum, 1 sensor of 120 Ω maximum
Calibration methods Actual load calibration
Digital span (calibration without a mass)
Measurement range Zero calibration range : -35 to +35 mV (-7 to 7 mV/V)
Span calibration range : -35 to +35 mV (-7 to 7 mV/V)
Minimum guaranteed input sensitivity : 0.4 µ V/d
Minimum display input sensitivity : No input sensitivity limit
Maximum display ±9999
Linearity 0.01% F.S. ±1 digit or less
A/D conversion 10 times/sec.
Temperature characteristics Zero : ±0.2 µ V/°C (Typ.)
Span : ±30 ppm/°C (Typ.)
Display panel 4 red 7-segment LEDs with a height of 14 mm, along with a minus sign and 5 status indicators
CADCAD File (DXF) Some characters may be garbled as the original CAD data is in Japanese.